MODEL AVOT-600
|
SPECIFICATIONS |
INPUT
|
120V 60 Hz, with
2.75 m. separable line cord, 3 conductor (240V 50 Hz input also available). |
OUTPUT
|
0 ... 60kV A.C. at
500 VA, centre tap grounded.
Maximum voltage from ground 30 kV. Crest factor better than 5%. |
CONTROL
|
Output continuously
adjustable by motor-driven control with 3 preselected speeds - 500V/sec. as per
ASTM-D1816, 2000V/sec as per IEC Publication 156, and 3000V/sec. as per ASTM-D877.
Interlock prevents starting except at zero voltage.
Cut out shuts off high voltage at breakdown, but retains failure voltage on the 115 mm
rugged meter. |
OIL CUPS
|
Interchangeable oil
cups sit in cradle contacts inside a test chamber with safety H.V. interlock and hinged
acrylic lid. The cups are demountable for polishing. A gap gauge to check spacing is
mounted on each cell. A plastic tray catches oil spills.
Oil Cup Model OC-77 meets the requirements of ASTM-D877.
Oil Cup Model OC-16 meets the requirements of ASTM-D1816, IEC 156 and VDE.
Includes an electric stirrer. |
CABINET
|
Sturdy steel
cabinet, with baked epoxy blue finish, with slip hinged lid, and comfortable heavy duty
carrying handles. |
SAFETY FEATURES
|
Line switch, line
fuse and pilot lamp. Safety interlock switch on sample cage. Automatic cut out shuts off
on breakdown. |
SIZE
|
38 x 38 x 30 cm |
WEIGHT
|
27 kg |